Background

The MiSeq i100 Series instruments are configured to automatically connect to a default set of Network Time Protocol (NTP) servers and synchronize the time settings on the instrument. Time synchronization is required for normal instrument operation and connecting to Cloud services such as BaseSpace Sequencing Hub.

The MiSeq i100 Series is configured to connect to the public 2.pool.ntp.org NTP server pool for time synchronization by default. If network security protocols restrict internet access, Illumina recommends adding this domain to the network security allow list to ensure access.

See the following instructions to configure specific NTP servers for either internet or local network time synchronization.

Resolution Actions

  1. Login to the MiSeq i100 Series Control Software as an Administrator user.

  2. Select the dropdown menu in the top-left corner, then select Settings, then select Time Settings.

  3. Select the appropriate Time Zone based on instrument location.

  4. Input the server name or IP address for local or internet NTP servers in the Network Time Protocol (NTP) text box, using commas to separate multiple input values.

  5. Select Save.

  6. Once completed, the Control Software will automatically restart.

For any feedback or questions regarding this article (Illumina Knowledge Article #10045), contact Illumina Technical Support techsupport@illumina.com.