> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009998.md).

# Air Filter replacement on the MiSeq i100 Series

The air filter is single use and covers the fan in the back of the instrument. It ensures proper cooling and prevents debris from entering the system. The instrument is shipped with one air filter installed and one spare. Additional air filters are included with an active instrument service contract (filters are not sent automatically; rather, users can request them by sending an email to Illumina Technical Support, <techsupport@illumina.com>), or they can be purchased separately from Illumina.

MiSeq i100 Air Filter catalog number: **20116201**

Use the following instructions or reference [MiSeq i100 Series system guide](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/ReplaceAirFilter.htm) to replace an expired air filter every 6 months.

**Replace Air Filter**

1. Position the instrument to easily access the back of it.
2. On the back of the instrument, pull the top edge of the rear panel away from the instrument to access the air filter.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-56a9fbd9ff215b2521767a2a6f7dbf826eed8813%2Fimage1.png?alt=media\&token=2dae9a83-f6f1-4cae-a4d2-1cf618894a51)

3. Remove and discard the used air filter.
4. Insert the new air filter into the tray. Ensure the filter is inserted so that the filter tab faces outward and sits against the rear panel.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-720678862f5b19dbae7c587e03bf226778033079%2Fimage2.png?alt=media)

5. Close the rear panel.
6. Select **Reset filter expiry**.
7. Return the instrument to its original placement.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9998), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009998%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009998.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.