> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000001777.md).

# How to perform a System Check on the MiSeq

System checks are not required for the normal operation and maintenance of an Illumina sequencing platform. Rather, the system checks are used for troubleshooting purposes.

* For all system checks, if an instrument wash is due, perform the wash before starting a system check.
* Always power cycle before performing the system checks.
* System check results are located in the D:\DiagResults folder.

**Instruction Steps:**\
![](/files/iSKv0Uf5Cy0zzTbOPRfp)

\* The Tip/Tilt and Full Optics tests require a special beaded flow cell and must only be performed by an Illumina Field Service Engineer (FSE).

\*\* If performing a volume test on a system running MiSeq Control Software (MCS) 2.6 or earlier *only*, the sippers must be manually raised and lowered to check the PR2 position. When prompted, remove any droplets from the wash bottle sipper, as follows.

1. Slowly raise the sipper handle and check the wash bottle sipper for the presence of a large water droplet.
2. Slowly lower the sipper handle far enough into the water to allow the surface tension to remove the droplet.
3. Slowly raise the sipper handle and check the wash bottle sipper for the presence of a large water droplet.
4. Slowly lower the sipper handle completely, making sure that the sippers lower into the wash bottle and waste bottle.

**Additional Resources:**

* MiSeq Control Software v4.0 or above: Page 39 of the [MiSeq System Guide on Windows 10](https://support.illumina.com/sequencing/sequencing_instruments/miseq/documentation.html).
* MiSeq Control Software v3.1 or below: Page 42 of the [MiSeq System Guide for Windows 7 and MiSeq System Guide for MiSeq Reporter](https://support.illumina.com/sequencing/sequencing_instruments/miseq/documentation.html).
* If any system check fails, consult this article [Troubleshooting failed MiSeq system checks](/instrumentation/miseq-ruo/instrumentation-miseq-ruo-troubleshooting-list/000007050.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1777), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001777%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000001777.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.