> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000010276.md).

# Troubleshooting the iSeq 100 v3.2.0 "Firmware Connectivity Problem Detected" initialization error

**Background**

The iSeq 100 Control Software v3.2.0 was released on 12-OCT-2026. Updated systems require an additional step to ensure that the instrument is able to be shut down/power cycled properly when needed.

The default behavior when the instrument is shut down via the Windows menu is to enter a hibernation state. When this shut down method is used, and the instrument is restarted, the following initialization error may present:

"Firmware Connectivity problem detected"

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6ed0a8a27ba602b817df855cb4909bb2abd0fb73%2Fimage1.jpg?alt=media)

**Troubleshooting steps**

To avoid this, disable the hibernation feature following the steps below.

1. In the Windows search bar, type "cmd" then right-click on **Command Prompt** and select **Run As Administrator**.
2. Enter `powercfg /hibernate off then press enter.`
3. Enter `powercfg /a then press enter.`
4. Check the "Hibernate" value shows "Hibernation has not been enabled".

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-c0f65251784fc4b668d70c3083750fc94e5613ee%2Fimage2.jpg?alt=media)

5. Close the Command Prompt application.
6. In the Windows Search bar, enter "regedit". Select **Registry Editor**, right-click and select **Run as Administrator**.
7. Navigate to location: Computer\HKEY\_LOCAL\_MACHINE\SYSTEM\CurrentControlSet\Control\Power
8. Search for "HiberbootEnabled". If present in this location, continue with step 9.

* **Note**: If "HiberbootEnabled" is not present in the above location, navigate instead to Computer\HKEY\_LOCAL\_MACHINE\SYSTEM\CurrentControlSet\Control\SessionManager\Power, locate "HiberbootEnabled" here, and proceed with step 9.

9. Double-click on "HiberbootEnabled" entry. If its value is "1", change it to "0".
10. Exit Registry Editor.
11. Repeat Steps 6-10 and verify "HiberbootEnabled" value is "0".

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-42faaa5fc9ea592dc54ba005b9de125b5a570419%2Fimage3.jpg?alt=media)

12. Ensure all open windows are closed, then select Windows Start, then **Power**, then **Shut Down** to shutdown the system. Once the system has powered down, wait 2 minutes, then proceed to the next step.
13. Toggle the power switch on the back to OFF.
14. Leave the system **off** for a minimum of **5 minutes**.
15. Switch the power switch on the back to **ON**. When the power button on the left side of the instrument starts to pulse, press it to power up the system.

**Correct Shutdown Procedure**

Each subsequent time the instrument is shut down or power cycled, follow the steps in [How to Power Cycle the iSeq 100](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002289.md).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10276), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010276%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000010276.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.