For the complete documentation index, see llms.txt. This page is also available as Markdown.

Troubleshooting the iSeq 100 v3.2.0 "Firmware Connectivity Problem Detected" initialization error

Background

The iSeq 100 Control Software v3.2.0 was released on 12-OCT-2026. Updated systems require an additional step to ensure that the instrument is able to be shut down/power cycled properly when needed.

The default behavior when the instrument is shut down via the Windows menu is to enter a hibernation state. When this shut down method is used, and the instrument is restarted, the following initialization error may present:

"Firmware Connectivity problem detected"

Troubleshooting steps

To avoid this, disable the hibernation feature following the steps below.

  1. In the Windows search bar, type "cmd" then right-click on Command Prompt and select Run As Administrator.

  2. Enter powercfg /hibernate off then press enter.

  3. Enter powercfg /a then press enter.

  4. Check the "Hibernate" value shows "Hibernation has not been enabled".

  1. Close the Command Prompt application.

  2. In the Windows Search bar, enter "regedit". Select Registry Editor, right-click and select Run as Administrator.

  3. Navigate to location: Computer\HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\Power

  4. Search for "HiberbootEnabled". If present in this location, continue with step 9.

  • Note: If "HiberbootEnabled" is not present in the above location, navigate instead to Computer\HKEY_LOCAL_MACHINE\SYSTEM\CurrentControlSet\Control\SessionManager\Power, locate "HiberbootEnabled" here, and proceed with step 9.

  1. Double-click on "HiberbootEnabled" entry. If its value is "1", change it to "0".

  2. Exit Registry Editor.

  3. Repeat Steps 6-10 and verify "HiberbootEnabled" value is "0".

  1. Ensure all open windows are closed, then select Windows Start, then Power, then Shut Down to shutdown the system. Once the system has powered down, wait 2 minutes, then proceed to the next step.

  2. Toggle the power switch on the back to OFF.

  3. Leave the system off for a minimum of 5 minutes.

  4. Switch the power switch on the back to ON. When the power button on the left side of the instrument starts to pulse, press it to power up the system.

Correct Shutdown Procedure

Each subsequent time the instrument is shut down or power cycled, follow the steps in How to Power Cycle the iSeq 100.

For any feedback or questions regarding this article (Illumina Knowledge Article #10276), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?