# iSeq 100 unable to connect to BaseSpace due to Proxy server issues

#### Phenotype

Universal Copy Service can have issues navigating a user's Proxy systems. This manifests as Basepace not connecting (BaseSpace Hosting locations missing in the software settings or other connectivity or upload issues).

![](/files/vXJdhN6sWtHQ1HGA9254)

#### Resolution

Follow the instructions below from the system guide on how to *Connect to a Proxy Server*. These instructions should be done only with a user's IT representative or system administrator:

1. Minimize the control software.
2. From Windows Start, open the Run dialog box.
3. Enter `cmd`, and then select **OK**.
4. Enter the following command:\
   C:\windows\System32\bitsadmin.exe /Util /SetIEProxy LocalSystem Manual\_proxy http\://: NULL
5. Replace http\://: with your proxy server address and proxy port and `NULL` with any bypasses.
6. Press **Enter** to run the command.
7. Power cycle the instrument. See [Power cycling the iSeq 100](/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002289.md).

Reference information in iSeq 100 Sequencing System support documentation: [Network Setup](https://support-docs.illumina.com/IN/iSeq100/Content/IN/iSeq/NetworkSetup_fISQ.htm?Highlight=proxy%20server).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2264), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002289%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000002264.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.