> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000010194.md).

# Troubleshooting 'libminizip.so.1 not found' errors when trying to start TeamViewer QuickSupport on NextSeq 1000/2000

**Background**

The TeamViewer QuickSupport program requires specific packages to run properly. In the NextSeq 1000/2000 series instruments running Oracle Linux 9, these packages may be missing. This results in the following error when attempting to launch the TeamViewer QuickSupport program:

"The following required libraries seem to be missing: libminizip.so.1 => not found. Please find and install the corresponding packages."

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ded08ca4ea74d770d1d932800e2e884335181fcb%2Fimage1.jpg?alt=media)

**Resolution Actions**

1. **Minimize** the NextSeq 1000/2000 Series Control Software.
2. Select the **Activities** menu in the top-left corner and open the Terminal application.
3. Input the command `su - ilmnadmin` and press **Enter**, then input the `ilmnadmin` password as prompted and press **Enter**.
4. Input the command `dnf list installed | grep minizip` and press **Enter**. If no result is output, the `minizip1.2` package is missing.
5. Input the command `sudo dnf makecache` and press **Enter**.
   1. If prompted, input the `ilmnadmin` password and press **Enter**.
   2. If the command shows an error message indicating "Errors during downloading metadata for repository 'sublime-text'", then repeat command as `sudo dnf makecache --disablerepo=sublime-text` and press **Enter**.
6. If the command from Step #5 completes successfully, input the command `sudo dnf install -y minizip1.2` and press **Enter**.
   1. If the command from Step #5 presented an error connecting to the 'sublime-text' repository, input command as `sudo dnf install -y minizip1.2 --disablerepo=sublime-text` and press **Enter**.
7. Once the `minizip1.2` package is installed, relaunch TeamViewer QuickSupport.
8.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10194), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010194%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000010194.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.