> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000009026.md).

# How to troubleshoot “A component failed during sequencing” error during a Cloud Outage on the NovaSeq 6000

**NOTE:** NovaSeq 6000 Control Software v1.8.2 introduces a mitigation for this particular scenario. See the [NovaSeq Control Software v1.8.2 Release Notes](https://support.illumina.com/content/dam/illumina-support/documents/downloads/software/novaseq/200072595_00_NovaSeq_6000_RUO_v1_8_2_Software_Customer_Release_Notes.pdf) for more information.

**Background**

When starting a NovaSeq 6000 Sequencing run and BaseSpace/Proactive are unavailable, the error message "Component failed during sequencing" can be observed shortly after starting the sequencing run.

![](/files/gblD28r6xV2XNfo1VBKQ)

**Troubleshooting steps**

If there is no sequencing run in progress on the other side, evaluate the following to see if the same reagents can be reloaded.

1. Confirm that the reagents are unpierced and no run folder has been generated under `C:\ProgramData*\Illumina\NovaSeq\NovaSeqTemp\`.
   * The `ProgramData` folder is Hidden by default. If this folder is not present, select the **View** tab in File Explorer and then select the checkbox next to **Hidden Items**.
2. Check the [Illumina Cloud Products Status](https://status.illumina.com/) page to see if BaseSpace is currently experiencing a service incident in the relevant region.

If either 1 or 2 are not applicable, refer to the Knowledge Base article [How to troubleshoot “A component failed during sequencing” error messages on the NovaSeq 6000](/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000001265.md) for more information.

**Resolution actions**

If there is a BaseSpace incident that is affecting NovaSeq 6000 run setup, see the following steps as a temporary mitigation until the incident has been resolved.

**Important**: Illumina recommends starting Xp flowcells within 30 minutes of loading the samples onto the flow cell. If 30 minutes have already elapsed from loading the Xp samples onto the flow cell, contact [Technical Support](https://www.illumina.com/company/contact-us.html) for further assistance.

1. From the NovaSeq 6000 Control Software home screen, select the dropdown menu in the top-left corner, then select **Settings**.
2. On the Settings page, de-select any checkboxes for **BaseSpace Sequencing Hub** or **Illumina Cloud** options.
   * When BaseSpace Sequence Hub is not enabled, an **Output Folder** is required.
3. Additionally, de-select the checkbox next to '**Send Instrument Performance Data to Illumina**'
4. Select **Save**.
5. Restart the sequencing run setup using the same consumables from the previous run attempt.
6. After the BaseSpace incident has been resolved, these Cloud settings can be reverted back to previous selections to return instrument to the previous workflow.

   ![](/files/kzYYex1lv4I218O86TYU)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9026), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009026%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000009026.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.