> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000006633.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Troubleshooting MiSeq Read 2 low intensity and quality scores

On the MiSeq platform, paired-end turnaround is performed after the i5 index read and before the start of Read 2. Low intensities and quality scores from the start of Read 2 are typically caused by issues with paired-end turnaround or poor priming Read 2.

Issues with paired end turnaround can be caused by the following root causes:

* Instrument fluidics.
* Reagent cartridge consumable quality.

Issues with Read 2 priming can be caused by the following root causes:

* Instrument fluidics.
* Reagent cartridge consumable quality.
* Library design or custom primers.

**Diagnosing**

The Intensity and % ≥ Q30 Data by Cycle charts will show show drops at the start of Read 2.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7ba8de5cdb8d6f4a1f9dc7af2f53290eda8c9701%2Fimage1.jpg?alt=media\&token=ab670d49-e927-4bdc-b881-2a55b2bc5f63)

Figure 1: Low Intensities in the Data by Cycle plot at the start of Read 2.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-9cf7c5705fa85eafa6f8fd58015309241944696b%2Fimage2.jpg?alt=media\&token=8774586a-ccdc-4972-8f91-8bb43103bdf2)

Figure 2: Low quality scores in the Data by Cycle plot at the start of Read 2.

**Troubleshooting steps:**

1. Once the run completes, complete the instrument post-run wash.
2. Perform the 'Fluidics' section of the MiSeq System Checks using the flow cell from the affected sequencing run.
   * Instructions for the MiSeq System Checks are found in the [How to perform a System Check on the MiSeq](https://github.com/illumina-swi/illumina-knowledge/tree/master/docs/Instrumentation/MiniSeq/Reference_Material/000001777.md) Knowledge article.
3. If the 'Fluidics' System Checks are all pass, the MiSeq fluidics system is operating within expectation.
   * The root cause of the issue was potentially a consumable related issue, proceed with repeat sequencing run if confident in Library Preparation design, quality, and primers.
4. If the 'Fluidics' System Check fails, perform the following actions:
   1. Power cycle the instrument.
   2. Switch to a trusted flow cell from a previously successful run.
   3. Perform a Maintenance wash.
   4. Repeat the 'Fluidics' system checks.

For further information, search for Knowledge articles:

* [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md)
* [Wash best practices for the MiSeq](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001679.md)

If the Fluidics system checks are unable to pass after performing these troubleshooting actions, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6633), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006633%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000006633.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
