> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000003666.md).

# Troubleshooting blank BaseSpace login screen during run setup for Windows 10 instruments (MiSeq, MiniSeq, NextSeq 500/550 and Novaseq 6000)

During early stage of run setup in the Control Software (CS) for any of the [Windows 10 based instruments](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-faq-list/000006353.md), users may be presented with a blank white screen instead of the BaseSpace Sequence Hub (BSSH) login prompt. For similar issues in iSeq 100, see the Knowledge Base article [Blank BaseSpace login screen during run setup in control software on the iSeq 100](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000002611.md).

**Note:** This issue will only impact users who are trying to send run data to BSSH for Run Monitoring and/or Storage or Proactive for Instrument Health Data sharing. If not using BSSH, the user can still set up the run.

**Possible causes:**

* Time on instrument is incorrect (this is the most common cause).
* Ports for BSSH are being blocked.
* Universal Copy Service (UCS) is not running.
* IE cache not clear.

**Troubleshooting steps:**

1. Update time to the correct internet time.
   1. Search **Time** in Windows and select **Change the data and time**.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-b740f9970c95797da5c5c154ea992ce0ac682cbe%2Fimage1.png?alt=media\&token=e9b96bbc-b7b7-45bf-a5db-ebd910ab372e)

1. 2. Toggle **Set time automatically** and **Set time zone automatically** and confirm the time is correct.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-333b210bf033240c9cb1fae8ecceeb7e128a1387%2Fimage2.png?alt=media\&token=773ad2e3-c200-44a6-b04e-cd87bd93663c)

2. Check that hosting location is selected in CS settings. If hosting location is already set, it may be necessary to re-save the hosting location:
   1. Navigate to **Manage Instrument** from the CS home page.
   2. Select **System Configuration** from Manage Instrument page.
   3. Select **BaseSpace Sequence Hub Configuration** from System Configuration page.
   4. Select a different hosting location from the one currently saved from the Hosting Location drop down list and select Save.
   5. To re-save the original hosting location, return to BaseSpace Sequence Hub Configuration and select the previously set hosting location from the Hosting Location drop down list, then select Save.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-845636d0d08b1f555bd6e4d645de88eb0573e207%2Fimage3.png?alt=media)

3. If time difference is large, it may be necessary to regenerate a Local Run Manager certificate:
   1. Open a web browser (by default is Chromium).
   2. Access <https://localhost>
   3. Select **Tools** from the upper left corner.
   4. Select **System Settings**.
   5. Navigate to the **Certificates** tab.
   6. Download and install the Local Run Manager certificate.
4. Confirm that all ports and domains required for BSSH are whitelisted.
   1. See [Illumina Proactive Tech Note](https://www.illumina.com/downloads/illumina-proactive-technical-note-m-gl-01092.html)
5. Confirm UCS is running under Windows Task Manager (check the **Services** tab).
   1. If not running, close CS.
   2. In Task Manager, right click on UCS and Start the service.
   3. Re-open CS and restart run set up.
6. Clearing the Internet explorer (IE) cache.
   1. Close CS if it is running.
   2. Click the Internet Explorer icon on the desktop or search for it in the Start menu.
   3. Click the gear icon in the upper-right corner of the browser window.
   4. From the Tools dropdown, select Internet Options.
   5. In the Internet Options window, ensure the General tab is selected (it is the default first tab).
   6. Under the Browsing history section, click the "Delete…" button.
   7. In the Delete Browsing History window, check Temporary Internet files and website files. Optionally, select Cookies, History, and other items.
   8. Re-open CS and restart run set up.

Contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) for additional assistance if needed.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3666), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003666%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000003666.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.