> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001808.md).

# How to restart Universal Copy Service when a run is stuck as uploading to BaseSpace Sequence Hub

The following are instructions on how to restart Universal Copy Service (UCS) when a run has completed successfully on a Windows-based sequencing instrument but the run is stuck or stalled in an "Uploading" state in BaseSpace Sequence Hub (BSSH).

* These instructions are relevant to Windows Instruments that run Universal Copy Service only.
* Restarting Universal Copy Service can restart the copy process to BaseSpace.
* This can be done by performing a power cycle on the instrument or following the instructions provided below.

**To Manually Stop and Start Universal Copy Service:**

1. Close the Instrument Control Software.
2. Open Task Manager as "administrator" by searching for Task Manager in the Windows Search bar, right clicking on "Task Manager", and selecting "Run as administrator"
   1. If the instrument is logged in as a user account, it will prompt to login with the admin credentials.
3. Stop Universal Copy Service in Task Manager by right clicking on UCS (can sometimes be found as ucs.exe or universal-copy-service) and selecting "End task"
4. Restart Universal Copy Service in one of the following ways:
   1. Power Cycle the instrument.
   2. For Windows Benchtop instruments (iSeq 100, MiniSeq, MiSeq, and NextSeq 500/550), navigate to the "Services" tab of Task Manager (with it still open as an administrator), Right click on Universal Copy Service, and select "Start"
   3. For NovaSeq 6000, opening the control software will start UCS or navigate to C:\Program Files\Illumina\UniversalCopyService then select the ucs.exe (sometimes listed as universal-copy-service.exe) executable to restart the service.
5. When Universal Copy Service is restarted it can restart the upload process to BSSH. When all files have completed upload, the run state will change from 'Uploading' to 'Pending Analysis.'

For reference, the following table reviews which sequencers and software versions use UCS and which use BaseSpace Broker.

**Table 1** - Universal Copy Service and BaseSpace Broker version based on Instrument Control Software version.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-087655dabff0b26b9a395d281c023d3aee3ca322%2Fimage1.jpg?alt=media)

To Restart BaseSpace Broker for upload to BaseSpace please reference the Knowledge Article: [Instructions on how to restart run upload to BaseSpace using BaseSpace Broker](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/Instructions-on-how-to-restart-BaseSpace-Broker/README.md).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1808), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001808%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001808.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.