> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001808.md).

# How to restart Universal Copy Service when a run is stuck as uploading to BaseSpace Sequence Hub

The following are instructions on how to restart Universal Copy Service (UCS) when a run has completed successfully on a Windows-based sequencing instrument but the run is stuck or stalled in an "Uploading" state in BaseSpace Sequence Hub (BSSH).

* These instructions are relevant to Windows Instruments that run Universal Copy Service only.
* Restarting Universal Copy Service can restart the copy process to BaseSpace.
* This can be done by performing a power cycle on the instrument or following the instructions provided below.

**To Manually Stop and Start Universal Copy Service:**

1. Close the Instrument Control Software.
2. Open Task Manager as "administrator" by searching for Task Manager in the Windows Search bar, right clicking on "Task Manager", and selecting "Run as administrator"
   1. If the instrument is logged in as a user account, it will prompt to login with the admin credentials.
3. Stop Universal Copy Service in Task Manager by right clicking on UCS (can sometimes be found as ucs.exe or universal-copy-service) and selecting "End task"
4. Restart Universal Copy Service in one of the following ways:
   1. Power Cycle the instrument.
   2. For Windows Benchtop instruments (iSeq 100, MiniSeq, MiSeq, and NextSeq 500/550), navigate to the "Services" tab of Task Manager (with it still open as an administrator), Right click on Universal Copy Service, and select "Start"
   3. For NovaSeq 6000, opening the control software will start UCS or navigate to C:\Program Files\Illumina\UniversalCopyService then select the ucs.exe (sometimes listed as universal-copy-service.exe) executable to restart the service.
5. When Universal Copy Service is restarted it can restart the upload process to BSSH. When all files have completed upload, the run state will change from 'Uploading' to 'Pending Analysis.'

For reference, the following table reviews which sequencers and software versions use UCS and which use BaseSpace Broker.

**Table 1** - Universal Copy Service and BaseSpace Broker version based on Instrument Control Software version.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-087655dabff0b26b9a395d281c023d3aee3ca322%2Fimage1.jpg?alt=media)

To Restart BaseSpace Broker for upload to BaseSpace please reference the Knowledge Article: [Instructions on how to restart run upload to BaseSpace using BaseSpace Broker](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/Instructions-on-how-to-restart-BaseSpace-Broker/README.md).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1808), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001808%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000001808.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.