> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000010190.md).

# Run Setup using the MiSeq i100 Series 100 cycle kit Turbo Custom Recipe

**Background**

The MiSeq i100 Series 100-cycle kit Turbo Custom Recipe supports an accelerated workflow for pathogen detection and identification, enabling 100 cycle single-end, dual-index runs to complete in 2.5-3 hours—compared to 4-5 hours for the standard MiSeq i100 Series recipes.

For more detailed information on the MiSeq i100 Turbo Custom Recipes with the Respiratory Pathogen ID/AMR Enrichment Panel, see the [Faster time to answer for infectious disease identification](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/miseq-i100-turbo-recipe-resp-path-tech-note-m-gl-04204/miseq-i100-turbo-recipe-respiratory-pathogens-tech-note-m-gl-04204.pdf) Tech Note.

The MiSeq i100 Series Turbo Custom Recipes are available by contacting [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

**Important Notes**

* The Turbo Custom Recipes are only compatible with **100-cycle kits**, though can be used with the 25M, 50M, or 100M flow cell configurations.
* The Turbo Custom Recipes have been tested for use with the [**Respiratory Pathogen ID/AMR Enrichment Panel (RPIP) Kit**](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/respiratory-pathogen-id-panel.html#tabgroup-1-tab2) (Illumina, Catalog 20047050)
* Users can use Turbo Custom Recipes with other Workflows or Applications, though Quality and Yield might be reduced and require optimization by the user.
* Run times with the Turbo Custom Recipes are optimized for **single**-**end**, **dual**-**index** configurations.
* Turbo Custom Recipe runs must be setup in Index First mode if index reads are performed.
* Paired-end runs are supported and will **add about 1 hour** to the total run time.
* The Turbo Custom Recipes require [**Control Software version 1.1.0**](https://support.illumina.com/sequencing/sequencing_instruments/miseq-i100-plus/downloads.html) or higher.
* The first few runs using the Turbo Custom Recipe will show projected run times consistent with Standard Recipe run times until the instrument completes several runs using the Turbo Custom Recipes.

**Run Setup Instructions**

1. Transfer the Turbo Custom Recipe files to a folder on a configured **External Storage** location, either network storage or USB drive, that is connected to the MiSeq i100 Series Instrument.

* Instructions for configuring **External Storage** are available in the [MiSeq i100 Series Product Documentation](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/SysOutFolder.htm)

2. In the MiSeq i100 Series Control Software, select **Start**, then proceed with Run Setup using Run Planning, Sample Sheet, or Manual (BCL Only).

* For additional details on setting up a Sequencing Run, see the [Plan a Sequencing Run](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/PlanRun.htm) section of the MiSeq i100 Series Product Documentation.

3. In the **Review run** user interface (UI), locate the Custom recipe section and click **Select**... (Figure 1).
4. In the **Select recipe file** UI, select the custom recipe file for the specific Flow Cell configuration used (Figure 2). Custom Recipe files are specific to the Flow Cell used, failure to select the correct Custom Recipe may result in Run failure.

* 25M/100cycle kit requires: 25M\_recipe-100-IF\_Turbo\_25M.xml
* 50M/100cycle kit requires: 50M\_recipe-100-IF\_Turbo\_50M.xml
* 100M/100cycle kit requires: 100M\_recipe-100-IF\_Turbo\_100M.xml

5. In the **Review run** UI, verify the Custom recipe is correct, then select **Load consumables** to proceed with Run Setup (Figure 3).

   ![](/files/wvYpJNCRjQKPTwWv8WsS)

**Figure 1**. The custom recipe selection link on the Review run page.

![](/files/sBDsOKGeX2wUkZQ33SEj)

**Figure 2**. Selecting the appropriate custom recipe file based on the Flow Cell configuration used.

![](/files/BeS0XmFdN0nKlMwNVlYT)

**Figure 3**. Verifying the correct Custom Recipe is selected, then select Load consumables to proceed.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10190), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010190%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000010190.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
