> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000010171.md).

# Using custom sequencing recipes on Illumina NovaSeq X Series instruments

**Background**

Custom sequencing recipes are sometimes used for specific sequencing applications; these recipes introduce changes to the standard imaging and/or chemistry steps performed by the sequencer. Illumina does not guarantee performance for runs performed with third party custom recipes.

**Run Setup using a custom recipes on Illumina NovaSeq X Series**

1. Transfer the Custom Recipe files to the `Downloads` directory on the NovaSeq X Series instrument.
   * Note: if using a USB device to transfer the custom recipe, for instruments running Control Software v1.3.1, see Knowledge Base article [Validated USB Storage Devices for NovaSeq X Series Control Software v1.3.1](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000009577.md) for approved USB external storage devices.
2. In the **Files** application, navigate to the `Downloads`directory.
3. If the files are compressed, right-click on the compressed Custom Recipe files and select **Extract Here**...
   1. Note that the uncompressed Custom Recipe files extract to a sub-directory within `Downloads`. Navigate to this sub-directory within `Downloads` in **Step 7**.
4. Within the NovaSeq X Series Control Software, select the **Start** button, then select **Sequencing**, and choose the appropriate **Side**(**s**) to begin Run Setup.
5. Toggle to **Manual** setup, then confirm that **Manual Run** (**generate BCLs Only**) is selected.
6. Input Run Name and cycle number information.
7. Under the **Review your run information** UI, click **Select**... under **Custom Recipe** (**Optional**) and browse to the uncompressed Custom Recipe file within the `Downloads` directory, then click the blue **Select** button.
8. Select **Load consumables** and proceed with run setup.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                         |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #10171), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000010171%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000010171.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.