> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000009867.md).

# PhiX Indexed Control (1000 Cycle) Product Information

**Background**

The PhiX control library is a well-balanced, high diversity library that can be used as a positive control during Sequencing Runs on Illumina platforms. However, due to an average insert size of 375 bp and total library size of 500 bp, the current PhiX v3 product is not compatible with Reads longer than \~350 cycles.

The new PhiX Indexed Control (1000 Cycle) formulation addresses this by increasing the insert length, enabling users to spike-in PhiX control when performing high cycle count Runs.

**Technical Details**

* Insert Size
* The PhiX Indexed Control (1000 Cycle) formulation has an average insert size of 700 basepairs (bp)
* The total library length is approximately 825 bp on average
* Indexing Strategy
* The PhiX Indexed Control (1000 Cycle) library leverages a 10 bp, dual-index strategy and is a pool of 5 index combinations
* This provides additional diversity and color balance in Index Reads when spiked into low-plexity library pools
* The index combinations do not overlap with any current Illumina or IDT index combinations
* For other Third Party Libraries, please review manufacturer's Library Preparation information for potential overlapping indexes before using the PhiX Indexed Control (1000 Cycle) library
* (**Optional**) Demultiplexing PhiX Indexed Control
* Adding the indexed PhiX libraries to the Sample Sheet for demultiplexing is not required.
* If the indexed PhiX libraries are not added to the Sample Sheet, the reads sort to the Undetermined FASTQ files, similar to un-indexed PhiX v3 Control.
* If required, add the PhiX Sample ID and Index combinations in **Table 1** to the Sample Sheet to generate FASTQ files for the individual PhiX libraries within the pool.

**Table 1**: PhiX Indexed Control (1000 Cycles) Index Sequences

| **PhiX Sample ID** | **i7 Bases for Sample Sheet** | **i5 Bases for Sample Sheet (Forward)** | **i5 Bases for Sample Sheet (Rev. Complement)** |
| ------------------ | ----------------------------- | --------------------------------------- | ----------------------------------------------- |
| `PhiX-Index1`      | `CGACCTAACG`                  | `ATACGCCGGC`                            | `GCCGGCGTAT`                                    |
| `PhiX-Index2`      | `TAGTTCGGTA`                  | `CTGTCACTTA`                            | `TAAGTGACAG`                                    |
| `PhiX-Index3`      | `ACCGGCCGTA`                  | `GGCGCCATTG`                            | `CAATGGCGCC`                                    |
| `PhiX-Index4`      | `AATCACCAGC`                  | `CAGTAATTAC`                            | `GTAATTACTG`                                    |
| `PhiX-Index5`      | `AGCGAATTAG`                  | `GGCGATCAGA`                            | `TCTGATCGCC`                                    |

* Loading Concentrations for the MiSeq i100 Series
  * See the [MiSeq i100 Product Documentation](https://support.illumina.com/sequencing/sequencing_instruments/miseq-i100-plus/documentation.html) for instructions on preparing PhiX spike-in controls.
  * See the [Maximizing performance on the MiSeq i100 Series Technical Note](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/miseq-i100-library-optimization-tech-note-m-gl-03322/miseq-i100-library-optimization-tech-note-m-gl-03322.pdf) for loading concentrations when setting up PhiX Control Run.
  * For other applications, the appropriate loading concentration and spike-in value must be empirically determined.
* Platform Compatibility
  * The PhiX Indexed Control (1000 Cycle) is compatible with all Illumina sequencing platforms.
  * Due to the larger total library length, the PhiX Indexed Control (1000 Cycle) may cluster less efficiently on some platforms, and optimal loading concentrations must be empirically determined.
  * When spiking-in the PhiX Indexed Control (1000 Cycle) into libraries that are less than 825 bp average size, increase the concentration of PhiX Indexed Control (1000 Cycle) library to offset the larger insert size.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9867), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009867%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000009867.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.