RTA4 Q Score Bin Ranges on the NovaSeq X Series

Background

The NovaSeq X Series utilizes Real Time Analysis v4 (RTA4) for base calling and quality scoring. There are three Q score bins used on the NovaSeq X Series, and these Q score calls are reported as an average for the range of called scores. See below for the relevant Q Score Bins and Ranges, depending on Control Software version.

Table 1. Control Software v1.4 (NovaSeq X Series Software 1.4.0 Customer Release Notes)

Bin

Q Score Range

2

NoCall, 0-2

9

3-17

24

18-29

41

30+

Table 2. Control Software v1.3 (NovaSeq X Series Software 1.3.0 Customer Release Notes)

Bin

Q Score Range

2

NoCall, 0-2

9

3-17

24

18-29

40

30+

Table 3. Control Software v1.2 and v1.2.2

Bin

Q Score Range

2

NoCall, 0-2

12

3-17

24

18-29

40

30+

For more information on changes to Q Score bins with Control Software v1.2, see the following Illumina article: NovaSeq X v1.2 software enables sequencing with 80% of bases >= Q40