> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002484.md).

# Washes on the NovaSeq 6000

**There are two types of washes on NovaSeq 6000.**

* Automatic post-run wash
* Manual Maintenance wash

**Automatic post-run wash:**

* Performed automatically after every completed sequencing run.
* The software automatically initiates an automatic post-run wash using the wash solution and NaOCl provided in the reagent cartridge.
* *Approximate time: 80 minutes.*

**Manual Maintenance wash:**

* A maintenance wash is required when:
  * There has not been a 4-lane run with a post-run wash within the last 14 days.
  * There has not been a maintenance wash within the last 14 days.
  * A run or post-run wash fails prematurely or is incomplete.
* *Approximate time: 80 minutes.*

**Note**: NovaSeq Control Software (NVCS) v1.4+ will have staggered start countdown timer, which also applies to being able to perform a wash. The NVCS staggered start time indicates the instrument's availability for starting a new run or a wash. A maintenance wash for a single side can only be started when the other side is either idle or performing SBS (sequencing chemistry) read cycles. For more information, refer to Staggered Start of Runs in the [NovaSeq 6000 Sequencing System Guide](https://support.illumina.com/downloads/novaseq-6000-system-guide-1000000019358.html).

**Required Reagents and Consumables for Maintenance Wash:**

1. 0.05% Tween 20 wash solution:

* To a 500 ml centrifuge bottle, add 0.2 ml 100% Tween 20 to 400 ml laboratory grade water.
* Invert to mix.
* Add wash/tween solution to the center well of the SBS wash cartridge. Fill to the fill line, which indicates the minimum required volume.
* The other reservoirs remain empty.

2. 0.25% NaOCl:

* To a 30 ml centrifuge tube, add 1 ml 5% NaOCl to 19 ml deionized water.
* Invert to mix.
* Load 5 ml of the prepared solution into position 17 of the cluster wash cartridge (the location is marked **Fill** and has an orange circle around it). All other reservoirs remain empty.

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e9e13a44d0ddedd72cf78ecd60a140da3648cd3a%2Fimage1.png?alt=media)

3. 4-lane wash flow cell (or a used 4-lane flow cell).
4. SBS wash cartridge.
5. Cluster wash cartridge.
6. Leave the used buffer cartridge from the previous run in place.

**Note:** A used buffer cartridge is required for maintenance washes to collect the wash liquid used to wash the lines. The buffer cartridge should be emptied or changed regularly. **If a used buffer cartridge is not available, contact** [**Illumina Technical Support**](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

**Steps to Start a Wash:**

1. Load the Wash Flow Cell.
2. Load the Wash SBS and cluster cartridges.
3. Empty Used Reagent Bottles.

* **Note**: This step is **extremely** important because the small bottle collects used reagents even if the system is configured to route used reagents externally.

4. Start the wash.

For more detailed instructions on how to perform maintenance wash, refer to [NovaSeq 6000 system guide](https://support.illumina.com/downloads/novaseq-6000-system-guide-1000000019358.html) (pages 61-65).

**Note**: After the wash is completed, leave the consumables/cartridges in place until the next run.

**Maintenance recommendations for wash cartridges:**

* Clean and invert wash cartridge when not in use to prevent microbial growth/contamination.
* The lid for the SBS wash cartridge can bend sippers if not positioned correctly. A run can be started without the lid, as it is just a splash guard.

**Maintenance recommendations for Wash Flow Cell:**

* The wash flow cell is also known as S3 Wash Flow Cell
* The lifespan of a wash flow cell depends on the gaskets for proper sealing of the vacuum. Each wash flow cell typically lasts for approximately 100 washes.
* A wash flow cell with worn-out gaskets will fail the RFID registration of the wash flow cell, and cannot be used.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2484), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002484%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000002484.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.