How to Power Cycle the NovaSeq 6000

Illumina recommends power cycling, if needed, during troubleshooting. Regular weekly power cycles are not required on the NovaSeq 6000.

To power cycle the NovaSeq 6000, use the following steps.

  1. From the main menu of the NovaSeq Control Software, select Shutdown Instrument.

  2. After the screen goes blank, switch the toggle power switch on the back of the instrument to the off position.

  3. Wait 1-2 minutes.

  4. Use the toggle switch to turn the instrument ON.

  5. Wait for the power button to start to pulse; this can take a few seconds.

  6. Use the power button on the right front side of the instrument to turn the instrument computer ON.

  7. Log in to Windows.

  8. Wait for Windows to load, then select the NovaSeq Control Software icon to initialize the instrument.

This information is also available in video at How To Power Cycle NovaSeq 6000 Video