> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001481.md).

# How to achieve more consistent cluster density on Illumina sequencing platforms

Achieving optimal cluster density is critical to high-quality sequencing on MiniSeq, MiSeq, and NextSeq 500/550. On non-patterned flow cells, cluster density has a significant impact on run performance, specifically data quality and total data output. While underclustering can maintain high data quality, it results in lower data output. Alternatively, overclustering can lead to run failure, poor run performance, lower Q30 scores, introduction of sequencing artifacts, and lower total data output. This article summarizes the resources and best practices to avoid underclustering and overclustering, and to achieve more consistent cluster densities.

* Quality check and accurate quantification of libraries is critical. If not performed, it is the most common cause of inconsistent cluster density. Always follow methods listed in the [Library quantification and quality control quick reference guide](/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001247.md) article to check Illumina libraries.
* Follow the latest instrument-specific guidelines to dilute and denature libraries:
  * [NextSeq 500/550 Denature and Dilute Protocol Generator](https://support-docs.illumina.com/IN/dnd-wizard/Content/dnd/landing.htm?ins=nextseq550)
  * [MiSeq Denature and Dilute Protocol Generator](https://support-docs.illumina.com/IN/dnd-wizard/Content/dnd/landing.htm?ins=miseq)
  * [MiniSeq System Denature and Dilute Libraries Guide](https://support.illumina.com/downloads/miniseq-denature-dilute-libraries-guide.html)
* Best practices for library denaturation:
  * The NaOH stock solution must have a pH >12.5 prior to any dilution. Check the pH of the stock solution before diluting. The recommended stock concentration is found in the appropriate instrument system guide or denature and dilution guide.
  * Always prepare freshly diluted NaOH for denaturing libraries at time of use.
  * To prevent small-volume pipetting errors from affecting the final NaOH concentration, prepare at least 1 ml of freshly diluted NaOH.
  * For best results, thaw sequencing reagents before denaturing and diluting libraries. For further instructions, see the system guide for the instrument and the article: [How to thaw and store sequencing reagents for optimal performance](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/How-to-thaw-and-store-sequencing-reagents-for-optimal-performance-support-bulletin/README.md)[.](https://support.illumina.com/bulletins/2016/07/how-to-thaw-and-store-sequencing-reagents-for-optimal-performance-.html)
  * Follow the denature and dilute libraries guide for the instrument.\
    \- For the MiSeq series, it is important that the final solution of NaOH is not greater than 1 mM NaOH after diluting with HT1. Greater than 1 mM NaOH inhibits template hybridization efficiency.\
    \- When denaturing libraries for NextSeq 500/550 and MiniSeq, use 200 mM Tris-HCl pH 7.0 to ensure that the NaOH is fully hydrolyzed in the final solution. As a result, template hybridization is not affected even when the final NaOH concentration is greater than 1 mM (see NextSeq 500/550 and MiniSeq denature and dilution guides, above).
* For highly structured or GC-rich libraries, cluster density consistency improves when the library is heat denatured before loading the pool onto the instrument. This method is optional for other libraries.
  * After NaOH denaturation of the sequencing library and dilution with HT1 to the final loading concentration, incubate the diluted library at 96Â°C for 2 minutes using a heat block.
  * After the heat incubation, invert the tube 1-2 times to mix.
  * Quickly move the library to an ice water bath for 5 minutes. The quick cooling step helps lock the library in its single stranded form.
  * Proceed immediately to cluster generation.
* Consider the average size of the library. Because the clustering process preferentially amplifies shorter libraries in a mixture of fragments, large libraries tend to cluster less efficiently than smaller libraries. Loading concentration must be optimized for each library type.
* Special consideration is needed when sequencing low-diversity libraries to ensure consistent cluster density and good data quality. See knowledge article: [What is nucleotide diversity and why is it important?](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/What-is-nucleotide-diversity-and-why-is-it-important-support-bulletin/README.md) For low diversity sequencing guidelines specific to NextSeq 500/550 and MiniSeq, refer to the knowledge article: [Best practices for low diversity sequencing on the NextSeq 500/550 and MiniSeq systems](https://github.com/illumina-swi/illumina-knowledge/tree/master/articles/Knowledge/Best-practices-for-low-diversity-sequencing-on-the-NextSeq-500-550-and-MiniSeq-systems-support-bulletin/README.md)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1481), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002882%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000001481.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.