> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general-library-prep/instrumentation-general-library-prep-reference_material-list/000008912.md).

# How to optimize library insert size representation on NovaSeq X Series instruments

**Background**

Shorter insert sizes can result in slight reductions in variant calling for single nucleotide variants (SNVs) and insertions/deletions (indels). The overall representation of insert sizes within a sequenced library pool is affected by several different factors. This article discusses these major factors with respect to sequencing on the NovaSeq X Series, and how users can optimize their workflow to achieve a representative range of insert sizes from within their library pool.

Insert size representation on the NovaSeq X Series is affected by the following parameters:

* Loading Concentration
* Insert Lengths within a Library
* Library Cleanup
* Library Type
* Library Pool

**Loading Concentration**

Loading concentrations above the optimal range can result in declining mean insert size for some libraries as shorter fragments preferentially cluster under these conditions. In this example, a TruSeq DNA PCR-free library was prepared targeting an insert size of 450-500 bp. The mean insert size when loaded at 75 pM on NovaSeq X Series 10B flow cell was 470 bp, while this decreased down to a mean of 450 bp when loaded at 110 pM.

In these cases, reducing the loading concentration can increase mean insert size and result in data that is more representative of the actual library insert size. In practice, insert size should be considered along with other factors such as percent passing filter (%PF), %Duplicates, and Coverage.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-25e82d5492d5c62a60920e6b1611df287705570d%2Fimage1.png?alt=media)

**Insert Lengths within a Library**

Wide range of insert size distribution within a pool might bias mean insert size towards short insert clustering. Increasing loading concentration influences preference for short inserts within a library pool, which can also lead to increased variation of index representation within the pool. This can make it more difficult to ensure each sample is obtaining the minimum number of reads required for analysis when loading concentrations are too high.

For example, the %Reads each sample within a pool obtains increases in spread (reflected in increasing pooling coefficient of variation - CV) as the loading concentration increases from 80 pM to 120 pM, with TruSeq DNA PCR-Free libraries that have a 30 bp insert size difference.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-be05cb88a0655e2160bbc74dba12260eb29348d6%2Fimage2.png?alt=media)

Individual samples within a pool that have shorter sizes will also get more reads, increasing index CV. For example, a TruSeq DNA PCR-Free library pool with 30 bp insert size difference, increasing the larger inserts in the pool by 20% results in significantly lower CV compared to the equal molar pooling condition. This requires every library to be run on a Bioanalyzer or TapeStation to determine average insert size.

**Library Cleanup**

Illumina recommends a library clean up step to reduce very short inserts and adapter dimers and minimize the impact to CV. Additional cleanup can help improve primary and secondary metrics by tightening insert size distribution. Insert size representation can be altered by applying additional bead cleanups to the final library pool, using modified bead ratios. For example, performing an additional cleanup of Illumina DNA PCR-Free libraires can increase average insert size.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-442deeff4ff2d5b3a6d9cc9c10c88d8975a9fd86%2Fimage3.png?alt=media)

See [Optimal variant calling with Illumina DNA PCR-Free Prep on the NovaSeq™ X Series](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/ilmn-dna-pcr-free-prep-novaseq-x-tech-note-m-gl-02388/ilmn-dna-pcr-free-prep-novaseq-x-tech-note-m-gl-02388.pdf) Tech Note for more details.

**Library Pool**

Mixing different libraries pools together on same lane show short insert clustering preference. Different library types can have different clustering efficiencies, even when they are similar in size. Additionally, shorter libraries cluster more efficiently and will obtain more reads relative to longer libraries. Therefore, mixing different library types in the same sequencing lane can have unpredictable outcomes.

* Illumina does not recommend mixing library types with insert sizes on the extreme ends, eg, combining miRNA libraries (\~150 bp total library size) with whole-genome sequencing libraries (\~600 bp total library size) in the same lane.
  * In such cases, shorter libraries typically cluster with higher efficiency, tilting the balance toward shorter inserts.
  * When mixing samples of varied sizes, consider if there is a chance to run into adapter sequence of the shorter libraries, as this may cause quality issue for the longer libraries from the same pool.
* Mixing libraries that should be sequenced with asymmetrical read lengths (like certain single cell RNA library types) with other libraries in the same lane is not recommended.
* Exercise caution while combining with Illumina Stranded RNA Prep libraries which require a dark cycle recipe applied to the entire flow cell for optimal performance. Dark cycle recipes for Illumina Stranded RNA libraries are only compatible with certain library types. For more information, see [Custom recipes for Illumina Stranded libraries on NovaSeq X Series](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000008394.md).

**No Significant Differences Across Flow Cell Types**

Internal studies show similar mean insert size for TruSeq DNA PCR-Free libraries sequenced across 10B and 25B flow cells when loaded at optimal concentrations. Illumina recommends titrating the loading concentration if the range of insert sizes is larger than expected when comparing similar library types on 10B and 25B Flow Cells.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6a7eb741308fe752443bc8d88cea032695d435b6%2Fimage4.png?alt=media)

**Library Type**

The impact of loading concentration changes is also affected by library type. Different library types have different compositions and size distributions, which can affect insert size as loading concentration is varied. For example, the insert size of Illumina Stranded Total RNA and Whole Exome libraries is less sensitive to loading concentration increases when compared to the TruSeq DNA PCR-Free data.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-d321a097db1d7ac8ed2c5fa6dae937919af11547%2Fimage5.png?alt=media)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8d7b3af08a483314f2284d5a893b743e2b7bebf1%2Fimage6.png?alt=media)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-149869622116df4df9270b68ee53a4c487ccb2e8%2Fimage7.png?alt=media)

Illumina has also seen a difference in insert size representation between the NovaSeq 6000 and NovaSeq X Series that is driven by library type. The largest differences between the NovaSeq 6000 and NovaSeq X Series occurs when insert size is greater than 200 bp:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-8a5900e94853f40aab0a3ba582ee3822cf79d37c%2Fimage8.png?alt=media)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-49cd0f3b06a21bddbc6296534df928a122f248de%2Fimage9.png?alt=media)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8912), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008912%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general-library-prep/instrumentation-general-library-prep-reference_material-list/000008912.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.