Troubleshooting "Unable to update index information of the Library" on the NovaSeq X Series
Background
Users may encounter the following error on the NovaSeq X Series when requeuing the analysis using the local DRAGEN workflows on the instrument with an edited v2 sample sheet.
"Unable to update index information of the Library because AllowIndexUpdates flag is false."
This error results from a discrepancy between the [Cloud_Data] and the [BCLConvert_Data] sections of the sample sheet. Users often edit the [BCLConvert_Data] section only, resulting in this error. While the error can be resolved by editing both sections, this is can be a difficult workaround.
It is recommended to instead create a new sample sheet in BaseSpace Run Planning, which will update both sections automatically. The user can import the original sample sheet into the BaseSpace Run Planning User Interface (UI), then make the appropriate edits and export the corrected sample sheet.
Use the new sample sheet when requeuing the analysis on the instrument using the local DRAGEN workflows. See How to requeue analysis on the NovaSeq X/X Plus for more information.